Performance, impact and experiences of using wearable devices for seizure detection in community-based settings: a mixed methods systematic review
Highlight box
Key findings
• Accelerometers performed well (sensitivity ≥80% and false alarm rate ≤1/day), at home among young patients with epilepsy in community-based settings.
• Patients felt uncomfortable and on spotlight when the device was too visible.
What is known and what is new?
• Wearable devices for seizure detection have high performance in inpatient settings. There is a gap of evidence on their performance, effect on health outcomes, and acceptability in outpatient and community-based settings.
• This systematic review provides evidence that wearable devices for seizure detection, especially accelerometers, perform well in community-based settings.
What is the implication, and what should change now?
• Furthers research were needed to examine the effect of wearable devices on health outcomes.
• Wearable devices’ developers must pay attention to their usability.
Introduction
Background
More than 50 million people live with epilepsy worldwide (1), with 30% presenting resistance to antiseizure medications (2). Seizures are associated with an increased risk of morbidity (e.g., burns, falls, dislocations, fractures and cerebral sequalae etc.) and mortality (e.g., sudden unexpected death in epilepsy) (3,4). Theses consequences are exacerbated by uncontrolled seizures and unsupervised environments, especially in home and in workplace settings (5).
Methods based on seizure self-reporting by patients or caregivers are often unreliable to optimize epilepsy management (6-10). Wearable devices for seizure detection (WDD) are mobile technological devices gathering biophysical signals during seizures to trigger alarms for patients or caregivers (11). These devices could improve safety, clinical management, self-management and quality of life (QoL) for patients with epilepsy (PWE). To achieve these outcomes, seizure detection devices should be accurate. Also, they should be acceptable and efficient for PWE (12).
Rationale and knowledge gap
There is some evidence that WDD can accurately identify generalized and focal-to-bilateral tonic-clonic seizures with high sensitivity and low false alarm rate (13-17). However, most diagnostic accuracy studies were conducted in inpatient settings (epilepsy monitoring units), limiting the safe transferability of results to outpatient settings (home and workplace). Research orientations from previous reviews suggest a focus on the performance assessment in outpatient settings (12,15,17). Previous reviews also lack report of WDD acceptability for PWE and caregivers, and their impact on health outcomes such as QoL, clinical management and self-management (12,14).
Objective
This systematic review aims to report the scientific evidence on: (I) performance of WDD to detect seizures in community-based settings (home, workplace and residential care setting); (II) impact of using WDD on health-related outcomes (QoL, seizure management, medication adherence and mortality), and (III) patient experience of using WDD. This review addressed outcomes that could inform evidence-based healthcare on epilepsy management at home for PWE and their informal caregivers, nurses, medical practitioners and policy makers according the FAME (Feasibility, Appropriateness, Meaningfulness and Effectiveness) scale proposed by Pearson et al. (18). We present this article in accordance with the PRISMA reporting checklist (available at https://mhealth.amegroups.com/article/view/10.21037/mhealth-24-7/rc) (19).
Methods
This systematic review follows the mixed methods systematic review approach (20,21). This protocol was registered on PROSPERO (Prospective Register of Systematic Reviews) before the selection step (CRD42020129787).
Eligibility criteria
Population
All studies about PWE were included in this review. Patients with acute seizures (i.e., related to an acute brain insult such as encephalitis, toxic-metabolic encephalopathies etc.) were excluded.
Intervention
WDD was the targeted intervention. The types of WDD were those already available on the market and validated in inpatient settings [accelerometer, electrodermal sensor, electrocardiogram (ECG), electroencephalogram (EEG), electromyogram (EMG), photoplethysmography, sound detection sensors, under-mattress device, blood oxygenation sensors, and multimodal sensors] (13-15).
Comparator
Patients in comparator group could receive inpatient or outpatient conventional video/EEG or usual care.
Outcomes
Three domains of outcomes were considered in this review: (I) performance of the device (sensitivity and false alarm rates); (II) impact of the device on health outcomes (QoL, clinical and self-management); and (III) patient experience about the device (perception, opinion and other experiences related to the use of WDD).
Type of studies
All empirical studies (quantitative, qualitative, mixed methods) performed in community-based settings were considered in this review. Studies from database creation up to 31 December 2022 were included. Only studies published in English were included. Editorials, comments, letters to the editor and technical memo were excluded.
Information sources
We developed a search strategy with a librarian with expertise in systematic reviews, for PubMed and Embase (OVID), Web of Science and Google Scholar. Articles published in the last 10 years (January 2013 to December 2022) were considered. The complete search strategy is available in Supplementary file (Appendix 1). We checked the references of eligible studies to retrieve studies that escaped our search strategy. Articles included in other systematic reviews published until 31 December 2022 were also retrieved as other sources of information.
Selection process
All references from databases and secondary sources were imported in Covidence (22), to manage the selection of relevant studies and the data extraction. Two independent reviewers (E.A. and J.M.W.S.) were involved in title and abstract assessment to apply the eligibility criteria on all potential sources. All sources retained were then independently evaluated using the full text to apply the eligibility criteria. When there was discrepancy between the two reviewers, a senior reviewer (M.S. or M.P.G.) made final eligibility decision.
Data extraction process
We developed a data extraction form and pretested it using two sources to ensure reliability. Two reviewers extracted the data using the Covidence online tool (22), and a senior reviewer validated all extractions.
We extracted the following information about studies: year, author, country, design, setting, method, and follow-up. Characteristics of patients were: age (mean or median with standard deviation or interquartile range) and sex (percentage of female). Clinical information collected included: type of seizures and number of seizures. Wearable device characteristics were: type of device, type of sensor, features of the device (place on body or home, remote alarm or monitoring, mode of signal transmission, battery autonomy, and mobility), and performance of the WDD (sensitivity and false alarm rate). We extracted information about health outcomes, including QoL, medication adherence, health professional or patient monitoring of seizures, and death. We operationalized the effects in a qualitative form: yes, no, or unclear. We also extracted information about the value, satisfaction, comfort, visibility, intrusiveness, and perceived utility of the WDD.
Data synthesis
We synthesized all this information separately by question about the performance, effect, and experience in narrative form.
Study risk of bias assessment
Studies on WDD performance (diagnostic test accuracy studies) were evaluated by using QUADAS-C (Quality Assessment of Diagnostic Accuracy Studies-Comparative) (23). Other quantitative, qualitative, and mixed methods studies were evaluated by using the Mixed Methods Appraisal Tool (MMAT) (24).
Results
Overall, 10 studies met the inclusion criteria and were used in the synthesis analysis. The flow diagram describes the selection process (Figure 1). From the 10 retained studies, six studies reported WDD performance in community-based settings (15,25-29). Two studies reported the effects of using WDD on health outcomes (mainly on QoL) in community-based settings (30,31). Finally, eight studies reported the experience of using the WDD in community-based settings (25-27,29-33) (Table 1).
Table 1
Authors | Country | Title | Setting |
---|---|---|---|
van Westrhenen 2021 (32) | The Netherlands | Parental experiences and perspectives on the value of seizure detection while caring for a child with epilepsy: a qualitative study | Home |
van Andel 2016 (15) | Belgium | Non-EEG based ambulatory seizure detection designed for home use: what is available and how will it influence epilepsy care | Home |
Thompson 2019 (30) | United States | Seizure detection watch improves quality of life for adolescents and their families | Home |
Olsen 2021 (33) | Denmark | Wearables in real life: a qualitative study of experiences of people with epilepsy who use home seizure monitoring devices | Home |
Meritam 2018 (25) | Denmark | User-based evaluation of applicability and usability of a wearable accelerometer device for detecting bilateral tonic-clonic seizures: a field study | Home and/or residential care |
Kjaer 2017 (26) | Denmark | Detection of paroxysms in long-term, single-channel EEG-monitoring of patients with typical absence seizures | Home |
Hadady 2023 (27) | Denmark | Real-world user experience with seizure detection wearable devices in the home environment | Home |
Dong 2021 (28) | China | Home-based detection of epileptic seizures using a bracelet with motor sensors | Home |
Borusiak 2016 (31) | Germany | A longitudinal, randomized, and prospective study of nocturnal monitoring in children and adolescents with epilepsy: effects on quality of life and sleep | Home |
Arends 2018 (29) | The Netherlands | Multimodal nocturnal seizure detection in a residential care setting: a long-term prospective trial | Residential care |
EEG, electroencephalogram.
A majority of studies were conducted in Europe [Belgium (15), Denmark (25,26,33), The Netherlands (29,32), and Germany (31)]. One study was conducted in the US (30) and another one in China (28). The majority of these studies were conducted in a home setting (15,25-28,30-33) (Table 1).
Characteristics of the devices are presented in Table 2. The majority of studies evaluated an accelerometer wearable device (15,25,27,28,30-33).
Table 2
Authors | Device | Place on body or home | Remote alarm or monitoring | Mobility | Battery autonomy | Mode of transmission |
---|---|---|---|---|---|---|
Arends 2018 (29) | Multimodal sensor bracelet | Upper arm | Unclear | Yes | NA | Transmission to a computer |
Borusiak 2016 (31) | EpiCare® (ACM) | Under the child’s mattress | Yes | No | NA | Transmission to a computer |
Borusiak 2016 (31) | Audio baby monitor | NA | NA | NA | NA | NA |
Dong 2021 (28) | Bracelet with ACM | Upper arm | Unclear | Yes | More than 24 hours | Storing data on a SD card |
Hadady 2023 (27) | NightWatch with a multimodal device based on ACM | Upper arm | NA | NA | NA | NA |
Hadady 2023 (27) | Empatica with a wristband with a multimodal seizure detection (ACM and electrodermal activity) | Upper arm | NA | NA | NA | NA |
Hadady 2023 (27) | EpiCare with a wrist-band with ACM-based seizure detection | Upper arm | NA | NA | NA | NA |
Kjaer 2017 (26) | Portable EEG device | Head | NA | NA | 24 hours | NA |
Meritam 2018 (25) | EpiCare (ACM) | Upper arm | Yes | Yes | NA | Transmission to portable control unit accessible to parents/caregivers or mobile phone |
Olsen 2021 (33) | Portable electroencephalography amplifier with two channels, an electrocardiography device | Sternum | Unclear | No | NA | NA |
van Andel 2016 (15) | Video, ACM, and radar-induced activity recording | The camera and radar are attached to a tripod that is placed close to the patient’s bed | Unclear | No | 10 hours | Transmission to a computer |
van Westrhenen 2021 (32) | NightWatch device | NA | NA | NA | NA | NA |
Thompson 2019 (30) | Smart Watch with 3D ACM | Wrist | Yes | Unclear | NA | Using Bluetooth to transmission signal to computer |
ACM, accelerometer; EEG, electroencephalogram; NA, not available; SD, secure digital.
Performance of WDD in a community-based setting
Studies that reported about the performance of WDD were cohort studies which compared the WDD with seizure diaries, carers’ report, or video. There were two cross-sectional studies which reported performance of WDD without comparison of the device with another tool to detect seizure. The follow up duration ranged from 15 to 450 days. The population were mostly young (children and adolescent) living at home or in residential care setting (intellectual disability) (Table 3). The most common type of seizures was tonic-clonic. Overall, the performance of all devices included in the review was good with high sensitivity (≥80%) and low false positive alarm rate (≤1/day) (Figure 2). These studies were of low quality according to the QUADAS-C (Table 4). This was mostly due to convenience sample that introduced selection bias and non-representativeness of the target population of the review. Some studies did not use comparator or reference standard to assess the sensitivity and false alarm rate (25,27). Also, it was unclear whether interpretation of the index test had introduced bias in most studies.
Table 3
Authors, year, country | Design/setting | Follow-up (days) | NS/NP | Population characteristics (age and sex) | Type of seizures | WDD | Sensitivity (%) | False positive alarm rate (per day) |
---|---|---|---|---|---|---|---|---|
van Andel 2016 (15); Belgium | Cohort study/home | 15 | 32/1 | 8 years; female (100%) | Tonic-clonic seizures | Accelerometer | 90.62 | 1.00 |
Meritam 2018 (25); Denmark | Cross sectional study/home and/or residential care | 450 | NA/71 | 27 years; female (45%) | Tonic-clonic seizures | Accelerometer | 85 | 0.10 |
Kjaer 2017 (26); Denmark | Cohort study/home | 30 | 593/6 | 10 years; female (83.3%) | Absence seizure | Portable EEG | 98.4 | 0.2 |
Hadady 2023 (27); Denmark | Cross sectional study/home | NA | NA/242 | 17 years; female (NA) | Tonic-clonic seizures | Accelerometer | 100 | 0.1 |
Dong 2021 (28); China | Cohort study/home | 117 | 114/5 | 29 years; female (40%) | Tonic-clonic seizures | Accelerometer | 75.91 | 0.13 |
Arends 2018 (29); Netherlands | Cohort study/residential care setting | 194 | 809/28 | 29.1 years; female (35.7%) | Tonic-clonic seizures | Accelerometer and photoplethysmography | 86 | 0.25 |
NS, number of seizures; NP, number of patients; NA, not available; WDD, wearable device for seizure detection; EEG, electroencephalogram.
Table 4
Authors | Patient selection | Index test | Reference standard | Flow and timing | |||||||
---|---|---|---|---|---|---|---|---|---|---|---|
Risk of bias | Concerns regarding applicability | Risk of bias | Concerns regarding applicability | Risk of bias | Concerns regarding applicability | Risk of bias | Concerns regarding applicability | ||||
van Andel 2016 (15) | Unclear | Unclear | High | High | Unclear | Unclear | Unclear | Unclear | |||
Meritam 2018 (25) | High | Unclear | High | High | High | High | High | High | |||
Kjaer 2017 (26) | Unclear | High | Unclear | Unclear | Unclear | Unclear | Unclear | Unclear | |||
Hadady 2023 (27) | Unclear | Unclear | High | High | High | High | High | High | |||
Dong 2021 (28) | Unclear | High | Unclear | Unclear | Unclear | Low | Unclear | Unclear | |||
Arends 2018 (29) | High | High | Low | Low | Unclear | Unclear | Unclear | Unclear |
Impact of WDD on health outcomes
Studies that addressed the impact of WDD on health outcomes (30,31) were about PWE’s QoL, parents’ anxiety and quality of sleep, and family support for child with epilepsy. There was no significant effect between groups and within groups on these outcomes in these studies (Table 5). The studies were of low methodological quality according to the MMAT (Table 6).
Table 5
Author/year/country | Design/setting | Follow-up (days) | Number of participants by group | Population characteristics (age and sex) | Type of seizures | WDD | Health outcomes | Effect (yes/no) |
---|---|---|---|---|---|---|---|---|
Thompson 2019 (30), United States | Mixed-method research/home (pre-post design) | 183 | One group (n=10) | 17.5; female =70% | Tonic-clonic seizure | Accelerometer | (I) Quality of life in Epilepsy for Adolescent (QOLIE AD-48). (II) Parent-child support. (III) Family life/leisure. (IV) Condition management. (V) Child autonomy. (VI) Child discipline | No for all outcomes |
Borusiak 2016 (31), Germany | Mixed-method research/home (pre-post and comparison design) | 210 | Accelerometer: n=13; control with nothing: n=16 | Accelerometer: 8.1 (3.6); female =55%. Control: 9.6 (4.5); female = 69.2% | NA | Accelerometer | (I) Parents’ fear of nightly seizures. (II) Change in parents’ QoL over time. (III) Frequency of co-sleeping. (IV) Change in parental quality of sleep | No for all outcomes |
Borusiak 2016 (31), Germany | Mixed-method research/home | 210 | Audio baby monitor: n=10; control with nothing: n=16 | Audio baby monitor: 9.5 (4.8); female =1%. Control with nothing: n=16 | NA | Audio baby monitor | (I) Parents’ fear of nightly seizures. (II) Change in parents’ QoL over time. (III) Frequency of co-sleeping. (IV) Change in parental quality of sleep | No for all outcomes |
NA, not available; WDD, wearable devices for seizure detection; QoL, quality of life.
Table 6
Authors | Are the participants representative of the target population? | Are measurements appropriate regarding both the outcome and intervention (or exposure)? | Are there complete outcome data? | Are the confounders accounted for in the design and analysis? | During the study period, is the intervention administered (or exposure occurred) as intended? |
---|---|---|---|---|---|
Thompson 2019 (30) | No | Yes | No | No | Can’t tell |
Borusiak 2016 (31) | No | Yes | No | No | Can’t tell |
Experience of using wearable device for seizure detection among PWE
Eight studies reported patient experience of using WDD in community-based settings (23-25,27-31). Documented attributes included perceived effectiveness, comfort, cost and usability.
PWEs and their caregivers perceived substantial effectiveness of using the WDD (25-27,29-31). Although there was high concern regarding the false alarm rate across studies, good sensitivity overcame this concern (27,30,31). False alarm rates were also reported due to the disconnection of device, difficulty to charge the device, phone intermediation, or other activities that could involve body motion (25,30). The other components of perceived effectiveness of the device were that users reported more freedom and safety (30), improvement in their QoL (25,27,30), improvement in seizure monitoring (27,29), and decrease in seizure related injuries (25,30).
Discomfort was reported by some users. This outcome was related to the visibility of the device. A highly visible device caused spotlight, vulnerability and social exclusion sensation for some users (33). Some users also expressed discomfort of wearing the device in public places (26,27,33). Some users reported the high workload of using WDD and life privacy invasion (29,32). Only one study reported that users found the WDD expensive (32).
Overall, studies reported ease of use of WDD, especially watch accelerometers (23,25,27,28). However, users reported that the device was not waterproof so they couldn’t wear it everywhere (25,30). Problems with charging or connection with the control unit were also reported (25,27,29,30). Overall, theses studies had good quality (Table 7).
Table 7
Authors | Is the qualitative approach appropriate to answer the research question? | Are the qualitative data collection methods adequate to address the research question? | Are the findings adequately derived from the data? | Is the interpretation of results sufficiently substantiated by data? | Is there coherence between qualitative data sources, collection, analysis and interpretation? |
---|---|---|---|---|---|
van Westrhenen 2021 (32) | Yes | Yes | Yes | Yes | Yes |
Thompson 2019 (30) | Yes | Yes | Yes | Yes | Yes |
Olsen 2021 (33) | Yes | Yes | Yes | Yes | Yes |
Meritam 2018 (25) | Yes | Yes | Yes | Yes | Yes |
Kjaer 2017 (26) | Yes | Yes | Can’t tell | Can’t tell | Can’t tell |
Hadady 2023 (27) | Yes | Yes | Yes | Yes | Yes |
Borusiak 2016 (31) | Yes | Yes | Yes | Yes | Yes |
Arends 2018 (29) | Yes | Yes | Yes | Yes | Yes |
Discussion
Key findings
The literature on the use of WDD in community-based settings is scarce despite the potential of this technology to improve the health and well-being of PWE and their caregivers. The WDD most used were accelerometers. Most studies included child and adolescent patients and focused on tonic-clonic seizures. The overall performance of using the WDD was high with over 80% of sensitivity and less than 1 false alarm per day. This good performance was reflected by users’ appreciation in the qualitative material. However, there were some discrepancies in the qualitative findings regarding the perceived false alarm rate which was deemed too high in some studies.
With respect to the second review question, quantitative studies did not report any effect of the use of WDD on QoL, parent child support, parents’ fear of seizures, parents’ quality of sleep, family life/leisure, condition management, and child autonomy. Qualitative studies reported that the use of WDD positively influenced QoL, condition management and seizure-related injuries. Finally, qualitative studies found that WDD were generally easy to use. However, users perceived more discomfort related to the visibility of the device.
Strengths and limitations
This review has some limitations. First, there is a possibility of publication bias because we searched publications in English only from two databases (PubMed, Google Scholar, Web of Science and Embase). Although, we could retrieve relevant papers from references of eligible studies and from other systematic reviews, some studies could have been missed. Second, the studies we synthetized in this review presented several risks of bias so the evidence presented in our review has a high risk of bias. Finally, we used a narrative approach to synthetize our findings, thus our subjectivity can affect the interpretation of the findings.
The main strength of this review is the systematic approach used to collect and synthetize data. The separate analysis of findings according to specific review questions is the preferred approach proposed for mixed methods systematic reviews (20).
Comparison with similar researches
The results of this mixed methods systematic review are consistent with those of previous reviews about the use of WDD. Previous reviews also reported high performance of WDD in in-patient settings and little concerns about false alarm rate (11,13-15). Previous reviews did not report about the impact of WDD on QoL, comfort, privacy, and impact of false alarms.
Explanations of findings
In this review, we reported no effect on QoL in quantitative studies but perceived effect on QoL in qualitative studies. This absence of effect on QoL in quantitative studies may be mainly due to small sample size (≤30) and perhaps short follow up time (≤365 days) used in these studies. Qualitative studies reported about the positive effect on seizure-related injuries and seizure management, but no quantitative evidence supported these findings. This may result from the difficulty of designing studies with sufficient power to detect these impacts. Our review also reported about comfort and highlighted that WDD, especially accelerometers, were suitable. Other devices which were more visible were less acceptable to users. This is consistent with a study that addressed PWE needs about WDD (34).
Implications and actions needed
The findings from this review suggest that WDD have a good performance to detect generalized tonic-clonic seizures and focal seizures in outpatient settings, but some concerns remain about the false alarm rate. More technological developments are needed to improve the performance of these devices, especially to reduce the false alarm rate. There was qualitative evidence that patients perceived improvement of their QoL and seizure management. Other quantitative studies are needed to support these findings. WDD were deemed comfortable by patients, except when they were too visible. It is thus important that WDD developers consider the visibility aspect to meet users’ needs. As such, involving users in the design of WDD is essential.
Conclusions
This mixed methods systematic review addressed the performance, impact on health outcomes, and patient experience of using WDD in outpatient settings. In general, WDD have a good sensitivity, but their false alarm rate is deemed too high. Qualitative studies reported positive impacts of WDD on QoL and seizure management, but quantitative studies reported no effect. WDD were also acceptable for patients, except when they were too much visible. More quantitative studies are needed to assess the impact of WDD on health outcomes.
Acknowledgments
We wish to thank Elisabeth Beauchamp-Chalifour (Centre intégré universitaire de santé et services sociaux de la Capitale-Nationale) and Tamara Herrera Fortin (Université de Montréal) for their involvement in the project and their work on the protocol for this systematic review.
Funding: This work was supported through the
Footnote
Reporting Checklist: The authors have completed the PRISMA reporting checklist. Available at https://mhealth.amegroups.com/article/view/10.21037/mhealth-24-7/rc
Peer Review File: Available at https://mhealth.amegroups.com/article/view/10.21037/mhealth-24-7/prf
Conflicts of Interest: All authors have completed the ICMJE uniform disclosure form (available at https://mhealth.amegroups.com/article/view/10.21037/mhealth-24-7/coif). M.P.G. serves as an unpaid editorial board member of mHealth from April 2024 to March 2026. The other authors have no conflicts of interest to declare.
Ethical Statement: The authors are accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.
Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/.
References
- Global, regional, and national burden of epilepsy, 1990-2016: a systematic analysis for the Global Burden of Disease Study 2016. Lancet Neurol 2019;18:357-75. [Crossref] [PubMed]
- World Health Organization. Epilepsy [Internet]. [July 26, 2023]. Available online: https://www.who.int/health-topics/epilepsy
- Sperling MR. The consequences of uncontrolled epilepsy. CNS Spectr 2004;9:98-101, 106-9. [Crossref] [PubMed]
- Yi Y, Zhong C, Wei-Wei H. The long-term neurodevelopmental outcomes of febrile seizures and underlying mechanisms. Front Cell Dev Biol 2023;11:1186050. [Crossref] [PubMed]
- Bertinat A, Kerr M, Cramer JA, et al. Living safely with epilepsy: a key learning review. Epileptic Disord 2020;22:364-80. [Crossref] [PubMed]
- Hoppe C, Poepel A, Elger CE. Epilepsy: accuracy of patient seizure counts. Arch Neurol 2007;64:1595-9. [Crossref] [PubMed]
- Blachut B, Hoppe C, Surges R, et al. Counting seizures: The primary outcome measure in epileptology from the patients' perspective. Seizure 2015;29:97-103. [Crossref] [PubMed]
- Elger CE, Hoppe C. Diagnostic challenges in epilepsy: seizure under-reporting and seizure detection. Lancet Neurol 2018;17:279-88. [Crossref] [PubMed]
- Tatum WO 4th, Winters L, Gieron M, et al. Outpatient seizure identification: results of 502 patients using computer-assisted ambulatory EEG. J Clin Neurophysiol 2001;18:14-9. [Crossref] [PubMed]
- Cook MJ, O'Brien TJ, Berkovic SF, et al. Prediction of seizure likelihood with a long-term, implanted seizure advisory system in patients with drug-resistant epilepsy: a first-in-man study. Lancet Neurol 2013;12:563-71. [Crossref] [PubMed]
- Li W, Wang G, Lei X, et al. Seizure detection based on wearable devices: A review of device, mechanism, and algorithm. Acta Neurol Scand 2022;146:723-31. [Crossref] [PubMed]
- Bruno E, Viana PF, Sperling MR, et al. Seizure detection at home: Do devices on the market match the needs of people living with epilepsy and their caregivers? Epilepsia 2020;61:S11-24. [Crossref] [PubMed]
- Beniczky S, Wiebe S, Jeppesen J, et al. Automated seizure detection using wearable devices: A clinical practice guideline of the International League Against Epilepsy and the International Federation of Clinical Neurophysiology. Clin Neurophysiol 2021;132:1173-84. [Crossref] [PubMed]
- Chen F, Chen I, Zafar M, et al. Seizures detection using multimodal signals: a scoping review. Physiol Meas 2022; [Crossref] [PubMed]
- van Andel J, Thijs RD, de Weerd A, et al. Non-EEG based ambulatory seizure detection designed for home use: What is available and how will it influence epilepsy care? Epilepsy Behav 2016;57:82-9. [Crossref] [PubMed]
- Casanovas Ortega M, Bruno E, Richardson MP. Electrodermal activity response during seizures: A systematic review and meta-analysis. Epilepsy Behav 2022;134:108864. [Crossref] [PubMed]
- Naganur V, Sivathamboo S, Chen Z, et al. Automated seizure detection with noninvasive wearable devices: A systematic review and meta-analysis. Epilepsia 2022;63:1930-41. [Crossref] [PubMed]
- Pearson A, Wiechula R, Court A, et al. The JBI model of evidence-based healthcare. Int J Evid Based Healthc 2005;3:207-15. [PubMed]
- Moher D, Liberati A, Tetzlaff J, et al. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. BMJ 2009;339:b2535. [Crossref] [PubMed]
- Pearson A, White H, Bath-Hextall F, et al. A mixed-methods approach to systematic reviews. Int J Evid Based Healthc 2015;13:121-31. [Crossref] [PubMed]
- Stern C, Lizarondo L, Carrier J, et al. Methodological guidance for the conduct of mixed methods systematic reviews. JBI Evid Synth 2020;18:2108-18. [Crossref] [PubMed]
- Veritas Health Innovation. Covidence systematic review software [Internet]. Melbourne, Australia; Available online: www.covidence.org
- Yang B, Mallett S, Takwoingi Y, et al. QUADAS-C: A Tool for Assessing Risk of Bias in Comparative Diagnostic Accuracy Studies. Ann Intern Med 2021;174:1592-9. [Crossref] [PubMed]
- Hong QN. Mixed Methods Appraisal Tool (MMAT) Version 2018 [Internet]. McGill University; [August 16, 2023]. Available online: http://mixedmethodsappraisaltoolpublic.pbworks.com/w/file/fetch/127916259/MMAT_2018_criteria-manual_2018-08-01_ENG.pdf
- Meritam P, Ryvlin P, Beniczky S. User-based evaluation of applicability and usability of a wearable accelerometer device for detecting bilateral tonic-clonic seizures: A field study. Epilepsia 2018;59:48-52. [Crossref] [PubMed]
- Kjaer TW, Sorensen HBD, Groenborg S, et al. Detection of Paroxysms in Long-Term, Single-Channel EEG-Monitoring of Patients with Typical Absence Seizures. IEEE J Transl Eng Health Med 2017;5:2000108. [Crossref] [PubMed]
- Hadady L, Klivényi P, Fabó D, et al. Real-world user experience with seizure detection wearable devices in the home environment. Epilepsia 2023;64:S72-7. [Crossref] [PubMed]
- Dong C, Chen L, Ye T, et al. Home-based Detection of Epileptic Seizures Using a Bracelet with Motor Sensors. In: 2021 10th International IEEE/EMBS Conference on Neural Engineering (NER); 2021:854-7.
- Arends J, Thijs RD, Gutter T, et al. Multimodal nocturnal seizure detection in a residential care setting: A long-term prospective trial. Neurology 2018;91:e2010-9. [Crossref] [PubMed]
- Thompson ME, Langer J, Kinfe M. Seizure detection watch improves quality of life for adolescents and their families. Epilepsy Behav 2019;98:188-94. [Crossref] [PubMed]
- Borusiak P, Bast T, Kluger G, et al. A longitudinal, randomized, and prospective study of nocturnal monitoring in children and adolescents with epilepsy: Effects on quality of life and sleep. Epilepsy Behav 2016;61:192-8. [Crossref] [PubMed]
- van Westrhenen A, de Lange WFM, Hagebeuk EEO, et al. Parental experiences and perspectives on the value of seizure detection while caring for a child with epilepsy: A qualitative study. Epilepsy Behav 2021; Epub ahead of print. [Crossref] [PubMed]
- Olsen LS, Nielsen JM, Simonÿ C, et al. Wearables in real life: A qualitative study of experiences of people with epilepsy who use home seizure monitoring devices. Epilepsy Behav 2021;125:108398. [Crossref] [PubMed]
- Herrera-Fortin T, Bou Assi E, Gagnon MP, et al. Seizure detection devices: A survey of needs and preferences of patients and caregivers. Epilepsy Behav 2021;114:107607. [Crossref] [PubMed]
Cite this article as: Sasseville M, Attisso E, Gagnon MP, Supper JMW, Ouellet S, Amil S, Assi EB, Nguyen DK. Performance, impact and experiences of using wearable devices for seizure detection in community-based settings: a mixed methods systematic review. mHealth 2024;10:27.